Polyethylene liner fracture in dual-mobility trapeziometacarpal joint prosthesis: Report of 17 cases
Published online on October 25, 2025
Abstract
Journal of Hand Surgery (European Volume), Volume 51, Issue 5, Page 639-642, May 2026.
We report 17 polyethylene liner fractures in 15 patients with dual mobility trapeziometacarpal joint prostheses who were primarily active males with high key pinch strength. Treatment involved head and liner exchange or trapeziectomy, depending on implant ...